    Calibration of glass electrode half-cells

    Agnes Heering, Ivo Leito

    Published: 2021-09-23 DOI: 10.17504/protocols.io.eq2lypyrmlx9/v2
    AI 解读
    浏览 1347

    Abstract

    The purpose of this document is give glass-electrode calibration procedure with aqueous buffers.

    Before start

    Turn on the thermostat and electrometer at least half an hour before starting the work.

    Steps

    Software

    1.

    Start Quick IV Measurement Software.

    Note
    Computer cannot go to sleep during measurements or the communication between computer and instruments is lost and data collection stops.

    2.

    On the left hand pick Function “Source & Sampling”.

    Position of the Function choice “Source & Sampling”.
    Position of the Function choice “Source & Sampling”.
    3.

    Right hand side click on tab “Channel Communication”. Click on “Search Channels”, which opens a new window. Select USB interface and search for the channel. After the search click on the channel and close the window.

    Position of the tabs “Channel Communication” and “Search Channels”.
    Position of the tabs “Channel Communication” and “Search Channels”.

    Note
    This must be done every time the program is started.

    4.

    Settings can be saved as QIVM file, which can be used for measurements. The saved file can be opened under tab “File”. The settings will remain the same between the measurements if the program is not closed.

    Program settings.
    Program settings.

    Note
    Range is "Fixed 20V" in case of EST-0601 calibration.

    Filling the cell

    5.

    25.0°C

    Equipment

    ValueLabel
    VistaShieldNAME
    Faraday cageTYPE
    GamryBRAND
    VistaShieldSKU
    http://www.gamry.com/LINK

    Note
    The side ports must be open.

    Note
    A beaker can also be used if temperature is controlled.

    6.

    Fill capillary with standard aqueous buffer with pH 7.00.50mL

    7.

    Insert glass electrode and a reference electrode.

    Equipment

    ValueLabel
    K401NAME
    Calomel reference electrodeTYPE
    RadiometerBRAND
    K401SKU
    7.1.

    Electrode is rinsed with water, gently dried, and then rinsed with solution to be measured.

    8.

    Connect electrodes to the instrument.

    Equipment

    ValueLabel
    B2987A Electrometer / High Resistance MeterNAME
    ElectrometerTYPE
    KeysightBRAND
    B2987ASKU
    8.1.

    Crocodile clips are as following:

    red is signal and connects to the glass electrode,

    black is shield and is connected to reference electrode and

    green is guard. Green stays unused.

    Note
    Cable goes through the port in Faraday cage. Electrometer is not in the Faraday cage.

    Equipment

    ValueLabel
    N1415A Triax to Alligator CableNAME
    CableTYPE
    Keysight TechnologiesBRAND
    N1415ASKU

    Measurement

    9.

    Start the measurement by clicking on the "Measure" button on the upper panel.

    10.

    Data collection. Point is taken at 10 s interval.

    Note
    Calibration duration must be the same as later used in pH measurements.

    11.

    Save data. Files are named as Date_GE_vs_ref_pHx.

    Note
    SAVE THE FILE BEFORE NEW MEASUREMENT! Otherwise, data is lost. Data is given in volts and seconds.

    11.1.

    Go to table view, right click on the table, choose “Export as CSV…” and save the results. The data can be automatically imported into excel file while measuring or after (both under “Excel Navigation”), but this causes constant shift in the comma position.

    Washing

    12.

    Remove the electrodes.

    13.

    Remove the solution.

    14.

    Rinse three times with plenty of water20mL

    15.

    Rinse once with acetone to ease the drying. 2mL

    16.

    Dry cell with compressed air.

    17.

    Repeat measurements with standard buffers pH 4 and pH 10.

    Data analysis

    18.

    Points from 30 min to 60 min are used for analysis.

    19.

    Take the average of the chosen data points.

    20.

    Plot Δ E vs pH.

    Calibration graph of a glass electrode half-cell.
    Calibration graph of a glass electrode half-cell.
    21.

    Obtain the slope and intercept of the glass electrode-half cell.