    Quant-iT™ PicoGreen® dsDNA Quantification

    Roey Angel, Eva Petrova

    Published: 2021-11-29 DOI: 10.17504/protocols.io.b2etqben
    AI 解读
    浏览 1594

    Abstract

    The following protocol is intended for the quantification of double-stranded DNA using Quant-iT™PicoGreen® dsDNA Assay Kit (ThermoFisher® dsDNA Assay Kit (ThermoFisher). This protocol is a simplified and condensed version of the full protocol from the manufacturer. The procedure described here is for 96 reactions. If samples are run in duplicates, then this should allow quantifying 40 samples.

    Before start

    This protocol is optimised for measuring an entire 96-well plate. It assumes that 16 wells will be used for measuring the standards and the blank samples (in duplicates) and 80 wells will be used for measuring unknown DNA samples (typically in duplicates).1. The protocol can be easily adjusted for a lower number of samples by reducing the volume of the working solutions of the reagents. Note though that enough TE should be retained for diluting the standard stock solution (490 or 680 µl), for potentially diluting the unknown samples, if their concentration is too high, and for accounting for pipetting errors. To fill the plate, 19.2 ml of TE is needed. So if only 40 wells are to be used for measuring unknown samples prepare about ml of TE buffer.

    1. The dynamic range of the assay is between 50 pg ml-1 to 1000 ng ml-1. This translates into DNA sample concentrations of 0.05-5 ng µl-1 and 1-200 ng µl-1 in the low-range and high-range assays, respectively. Samples with higher DNA concentration need to be diluted (e.g. in DNase-free water or TE buffer).
    2. Note that some compounds that can be present as DNA contaminations (e.g. salts, ethanol, detergents, proteins) are claimed by the manufacturer to not interfere with the measurement. Please refer to the full protocol for a list of these compounds and their effect on the measurement. Also, equimolar presence of ssDNA and RNA in the sample should have only minimal effect on the quantitation results.

    Attachments

    mp07581.pdf

    Steps

    Prepare reaction

    1.

    Take out all reagents from the fridge and bring them to room temperature.

    Take out the DNA samples from the freezer. DNA samples should be slowly thawed on ice.

    Note
    Quant-iT™PicoGreen® dsDNA reagent is dissolved in dimethylsulfoxide (DMSO), which freezes below 19 °C. The reagent must be completely thawed before using it by bringing it to room temperature. After the reagent thawed, it is advisable to briefly vortex the tube to make sure it is adequately mixed and to spin it down in a centrifuge or a mini centrifuge.

    Note
    Quant-iT™PicoGreen® dsDNA reagent is light sensitive and should be protected from light at all times.

    2.

    Prepare 22 ml 1X TE buffer by pipetting 1.1 ml of 20X TE buffer into 20.9 ml of nuclease-free water into a sterile and nuclease-free 50 ml tube.

    Mix by inverting the tube several times.

    1.1

    20.9

    3.

    For high-range quantification:

    Dilute the DNA-standard stock solution (λ DNA 100 ng µl-1) to a final concentration of 2 ng µl-1by mixing 10 µl λ DNA-standard stock solution with 490 µl 1X TE buffer.

    10

    490

    For low-range quantification:

    Prepare a 40-fold dilution of the 2 ng µl-1 DNA-standard work solution by mixing 5 μl of the 2 ng μl-1 DNA-standard work solution with 195 μl 1X TE buffer to yield a 0.05 ng µl-1 DNA-standard work solution.

    5µL

    195µL

    4.

    If needed, prepare a dilution of each sample in 1X TE buffer so that the reading will be within the dynamic range.

    Note
    It is advisable to run samples in duplicates for a more accurate quantification

    5.

    Prepare PicoGreen® work solution: 9950 µL 1X TE buffer + 50 µL PicoGreen®into a sterile and nucleic-acids free 50 ml tube. Mix and protect from light.

    9950

    50

    6.

    Prepare the following standard mixture in the first two columns of the black, sterile, 96-well plate:

    ABCD
    Assay versionDiluted DNA std. (µl)1X TE buffer (µl)Final DNA amount (ng)
    High-range (1-200 ng µl-1)1000200
    Use 2 ng μl-1 standard5050100
    109020
    1991
    01000
    Low-range (50 pg µl-1 - 5 ng µl-1)10005
    Use 0.05 ng μl-1 standard50502.5
    10900.5
    1990.05
    01000

    Equipment

    ValueLabel
    96-well microtiter plateNAME
    NuncBRAND
    265301SKU
    black, flat bottomSPECIFICATIONS
    7.

    Pipette 99 µl of 1X TE buffer in the remaining wells.

    99

    Note
    Tip: use a mechanical or electronic dispenser during this step and step no. 9 to speed up the work.

    Equipment

    ValueLabel
    Multipette E3NAME
    EppendorfBRAND
    4987000010SKU
    electronic dispenserSPECIFICATIONS
    8.

    Pipette 1 µl of the unknown DNA samples in the remaining wells.

    1µL

    Note
    Use either a diluted sample in case the concentration is expected to be higher than the dynamic range limit or larger volume in case the concentration is expected to be below the detection limit.

    9.

    Pipette 100 µl of the PicoGreen® work solution in each well, including the standard and unknown sample wells.

    100

    10.

    Protect the 96-well plate from light and incubate for 2-5 min at room temperature.

    0h 2m 0s

    Measure samples

    11.

    Place the plate in a plate reader and measure the fluorescence according to the following parameters:

    Excitation ~480 nm

    Emission ~520 nm

    Integration time 40 s

    Lag time 0 s

    Gain Optimal

    Number of flashes 10

    Calculated well highest standard

    Shaking 5 s

    Note
    It is also possible to set the gain to a fixed value (e.g. 100). If the fluorescence values of the standard drop over time this could indicate damage to the reagents or the DNA standard.

    Equipment

    ValueLabel
    Synergy 2NAME
    absorbance microplate readerTYPE
    BioTekBRAND
    Synergy2SKU
    12.

    Plot the measured fluorescent values of the standard samples against their known concentrations and fit a linear curve using linear regression. Make sure that the coefficient of determination (R2) is close to 1 (typically > 0.99). Calculate the DNA concentrations in the unknown samples using the slope and intercept parameters of the linear equation. Output values you obtained are in ng µl-1, assuming 1 µl of each sample was used.

    See attached example spreadsheet.

    Note
    Do not forget to account for any dilutions when calculating the concentration of the DNA in the unknown samples.
    PicoGreen_example.xlsx